Page 2
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
A representative of the manufacturer of the Navigation system should be on site for startup support. Required documents System requirement ARCADIS System Software VB 10C or higher Required parts TCP/ IP type crossover cable for a direct connection to the Navigation system (not included in the shipment)
Observe the following: - the product-specific safety information in these instructions, - the general safety information in the instructions TD00-000.860.01... and - the general safety information according to ARTD Part Siemens AG SPR2-310.814.02.02.02 ARCADIS Page 5 of 12 Medical Solutions 04.04...
Connect the TCP/IP crossover cable to the monitor trolley and the 2D Navigation sys- tem (see figure below). NOTE If the ARCADIS system is connected to the Navigation system via the clinical network, the connection may be affected by heavy net- work traffic.
Enter the password and click <ok>. • Click <Configuration>. • "Attached to Network" and "Dicom Networking" must be selected. • Click <Next>. Fig. 2: Configuration 1 Siemens AG SPR2-310.814.02.02.02 ARCADIS Page 7 of 12 Medical Solutions 04.04 CS SD 24...
Click "Find" to check whether the host is alive and ensure that the TCP/IP address of the Navigation system appears in the address field. • Click <Save>. • Click ">". Fig. 3: Configuration 2 ARCADIS SPR2-310.814.02.02.02 Siemens AG Page 8 of 12 04.04 CS SD 24 Medical Solutions...
Select "define new" under "General Node Properties" to edit the logical name "NAV2D_XXX" (XXX = logical name of Navigation system). NOTE If the ARCADIS system is also connected to a 3D Navigation sys- tem, the logical name should be "NAV3D_XXX" (XXX = logical name of Navigation system).
Installation and Settings Final steps • Perform a functional check. ARCADIS SPR2-310.814.02.02.02 Siemens AG Page 10 of 12 04.04 CS SD 24 Medical Solutions...
Page 11
Changes to previous version 3Changes to previous version None; initial publication Siemens AG SPR2-310.814.02.02.02 ARCADIS Page 11 of 12 Medical Solutions 04.04 CS SD 24...
Page 12
Changes to previous version ARCADIS SPR2-310.814.02.02.02 Siemens AG Page 12 of 12 04.04 CS SD 24 Medical Solutions...